Testonic® Chlorhexidine (CHX) Residue Test Kit is designed for fast and convenient measurement of chlorhexidine residues in water, rinse samples, and appropriately sampled surfaces following disinfection and antiseptic applications. Its drop-count titration method provides results with a factor of 1 drop = 5 ppm CHX.
The kit is suitable for monitoring equipment, tanks, process surfaces, and rinse water after disinfection. Its simple, instrument-free procedure provides numerical results in approximately 5 minutes.
Why Is Chlorhexidine Residue Monitoring Important?
Chlorhexidine (CHX) is a cationic active ingredient used in various disinfection and antiseptic applications. Monitoring remaining levels after treatment is important for evaluating rinsing operations and process conditions. Regular CHX testing makes it easier to monitor chemical residues and compare different sampling points.
Chlorhexidine Residue Testing in 3 Easy Steps
- Prepare the Sample: Place the water or rinse sample in the measurement vessel. For surface testing, follow the sampling procedure specified in the instructions.
- Perform the Test: Use the kit solutions as directed. Count the titrant drops while observing the color endpoint.
- Calculate the Result: Multiply the titrant drop count by 5 to determine the CHX concentration in ppm.
Kit Contents
- CHX Test Reagents
- Drop-Dispensed Titration Solution
- Measurement Accessories
- Instructions for Use
Testonic® Chlorhexidine (CHX) Residue Test Kit
Testonic® Chlorhexidine (CHX) Residue Test Kit is designed to measure chlorhexidine residues in water, rinse samples, and appropriately sampled surfaces following disinfection and antiseptic applications.
Its titration method provides numerical results with a factor of 1 drop = 5 ppm CHX under the specified test conditions. Residual chlorhexidine can be monitored in approximately 5 minutes without laboratory instruments.
Analytical Principle
The test is based on titrimetric evaluation of chlorhexidine in a sample using the kit reagents. During titration, the added drops are counted while observing the color endpoint.
According to the kit calibration, each titrant drop corresponds to 5 ppm CHX under the sample volume and test conditions specified in the instructions for use.
Why Is Chlorhexidine Residue Monitoring Important?
Chlorhexidine (CHX) is a cationic active ingredient used in various hygiene, disinfection, and antiseptic applications. When residual CHX must be monitored, visual cleanliness alone does not provide sufficient information.
Regular CHX measurements make it easier to monitor chemical residues following treatment or rinsing and compare the results with the facility's established acceptance criteria.
Applications
- Chlorhexidine-based disinfection processes
- Rinse water and suitable liquid process samples
- Residue monitoring on equipment and process surfaces
- Post-disinfection checks for tanks, containers, and equipment
- Veterinary hygiene applications
- Clinical and technical hygiene applications
- Routine chemical residue and quality control procedures
How to Use
- Prepare the Sample: Place the specified volume of water or rinse sample in the measurement vessel. For surface testing, follow the sampling and extraction procedure given in the instructions.
- Perform the Test: Use the kit reagents as directed. Count the titrant drops while observing the color endpoint.
- Calculate the Result: Multiply the titrant drop count by 5 to calculate the CHX concentration in ppm.
Technical Specifications
| Product Name | Testonic® Chlorhexidine (CHX) Residue Test Kit |
| Target Parameter | Chlorhexidine (CHX) |
| Drop Factor | 1 drop = 5 ppm CHX |
| Sample Type | Water, suitable liquid samples, and sampled surfaces |
| Test Method | Drop-count titration |
| Result Unit | ppm CHX |
| Analysis Time | Approximately 5 minutes |
| Test Capacity | Approximately 50 tests |
| Instrument Required | No |
| Storage Temperature | 8–24 °C |
Interpretation of Results
The CHX concentration is calculated from the number of titrant drops used.
CHX (ppm) = Number of Drops × 5
Example: If the endpoint is reached after 6 drops using the standard sample volume specified in the instructions:
6 × 5 = 30 ppm CHX
For liquid samples, the result represents the CHX concentration in the tested sample. For surface samples, the ppm result refers to the prepared extraction liquid and should not automatically be treated as the total residue amount on the surface.
Acceptability should be determined according to the intended application and the facility's established residue criteria.
Analytical Interferences and Measurement Limitations
- Other cationic active substances and unsuitable sample components may interfere with the titration result.
- Colored or turbid samples may make the endpoint difficult to identify.
- For surface testing, the sampling area, extraction method, and residue recovery may affect the result.
- The 1 drop = 5 ppm factor applies to the sample volume and conditions specified in the instructions.
- Failure to obtain a measurable result does not prove the absolute absence of CHX.
- The test does not measure microbiological cleanliness or disinfection effectiveness.
Related Residue and Hygiene Parameters
Other parameters that may be monitored after cleaning and disinfection include:
- QAC / Cationic Detergent Residue: Monitoring of quaternary ammonium compounds.
- Glutaraldehyde Residue: Monitoring of glutaraldehyde-based disinfectants.
- Hypochlorite Residue: Residue monitoring in hypochlorite-based applications.
- Active Oxygen Residue: Monitoring of hydrogen peroxide-based systems.
- Acid / Alkali Residue: Monitoring acidic or alkaline chemical effects after cleaning.
Although chlorhexidine is cationic, it is chemically different from quaternary ammonium compounds (QAC). CHX and QAC tests should not be used interchangeably.
Important Use Notes
- Follow the instructions for sample volume and titration procedure.
- For surface testing, use the sampling method specified in the instructions.
- When comparing surface results, use the same sampling area and extraction conditions.
- Use the test solutions with their original droppers.
- Store the reagents at 8–24 °C under the conditions specified for the product.